The HealthCXR Device is a software workflow tool designed to aid the clinical assessment of adult Chest X-Ray cases with features suggestive of pleural effusion in the medical care environment. The software automatically identifies suspected findings suggestive of pleural effusion on chest x-rays and notifies PACS/ workstation of the presence of the critical findings.
Clinical Workflow

For CalanticTM software version 1.0.0, no prior study will be examined.
*Part of Nanox AI's pneumothorax & pleural effusion triage solution.

AP or PA chest x-ray image
Image transformation of 0, 90, 180 or 270 degrees
A study with no more than 15 series
A series with no more than 5 images
If a pleural effusion is found in a chest x-ray, the software notifies PACS/workstation of the presence of this critical finding. In addition, the software provides visual output to the user intended to aid in the detection or ruling out of a pleural effussion.
Vendor
Nanox AI.
EU risk class and CE marking
HealthCXR Device has CE marking (CE2195) and risk class IIa.
Reimbursement status
Not reimbursed.
Contraindications
HealthCXR device should not be used to analyze scans of patients under 18 years old, scans which are not AP or PA chest x-ray, scans with severe missing image artifacts, scans missing parts of the chest cavity or scans that are unnaturally monochrome in nature (too black or too white). Non-standard scans may skew analysis results, and if such scans are sent to HealthCXR device, physicians should disregard results from the scans.
Patient Target Group
Patients 18 years and older.
Limitations
- The HealthCXR device is limited to the analysis of Chest X-ray scans.
- The HealthCXR device does not analyze scans of patients under 18 years of age.
- The determination of patient’s age is limited to reading the Birth Date and Study Date.
- DICOM attribute, or Patient’s Age DICOM attribute.
- The device does not analyze images that are not AP or PA Chest X-Rays.
- The device has limited capabilities to identify inadequate technical quality of x-ray images like image distortion, or presence of artifacts.
- The device has limited capability to handle image rotations that are not 0, 90, 180, or 270 degrees.
- The device does not process a study when the number of series exceeds 15.
- The device does not process a series when the number of images exceeds 5.


Discover HealthCXR Device
Now available as part of Calantic’s Thoracic Service Line. Contact us today to learn more.